Comprehensive Guide to Clinical Audiometric Masking
Principles of Auditory Cross-Hearing and Interaural Attenuation
Cross-hearing occurs when an auditory stimulus presented to a test ear (TE) travels across the head with sufficient intensity to be perceived by the cochlea of the non-test ear (NTE). The risk of cross-hearing is highest when the non-test ear possesses superior hearing sensitivity relative to the test ear.
Mechanics of Signal Crossover
When an air-conduction (AC) tone is presented at a high intensity level through headphones or insert earphones, the acoustic energy vibrates the transducer against the skull, inducing mechanical vibration of the cranial bones. This bone-conduction (BC) stimulation directly excites the fluids of both cochleae. Consequently, if the acoustic stimulus delivered to the TE, minus the energy lost across the head, meets or exceeds the bone-conduction threshold of the NTE, the signal will be detected by the non-test ear.
Interaural Attenuation Definitions and Baseline Values
Interaural Attenuation (IA) is defined as the loss of acoustic energy (expressed in decibels) as a sound signal travels from the test ear to the non-test ear.
Cross-Hearing vs. Interaural Attenuation: Cross-hearing refers to the clinical event wherein a test stimulus crosses the skull and is heard by the non-test ear's bone-conduction pathway. Interaural attenuation represents the quantitative loss of sound energy that occurs during this transmission.
The magnitude of interaural attenuation depends directly on the transducer type:
Air-Conduction via Supra-Aural Headphones:
Air-Conduction via Insert Earphones:
Bone-Conduction (BC Transducer):
Because the interaural attenuation for bone conduction is , a bone-conduction oscillator vibrates the entire skull, stimulating both cochleae equally regardless of placement location on the mastoid or forehead.



Criteria and Clinical Indications for Clinical Masking
Uncontrolled cross-hearing introduces significant clinical errors:
Test responses are recorded from the non-test ear rather than the test ear, generating a false or shadow audiogram.
True auditory thresholds of the test ear cannot be isolated or accurately established.
To eliminate cross-hearing, clinical masking is applied. Masking involves presenting a continuous noise signal to the non-test ear to temporarily raise its threshold, preventing it from detecting any crossover test stimulus delivered to the test ear.
Transducer Selection Advantages
Insert earphones are heavily favored over supra-aural headphones for pure-tone testing because:
Insert earphones provide a higher interaural attenuation ( vs. ), significantly reducing the frequency with which masking is required.
Insert earphones eliminate or minimize the Occlusion Effect (OE) encountered during bone-conduction testing when the non-test ear canal is covered.
Noise Types for Clinical Masking
Narrowband Noise (NBN): Used during pure-tone audiometry. NBN is centered around the specific test frequency. A narrow band of frequencies surrounding the pure tone provides maximum masking efficacy with minimal sound pressure level.
Speech Noise: Used during speech audiometry. Speech noise consists of broad white noise filtered to match the overall power spectrum of speech, centered around mid-frequency energy.
Criteria for Masking Pure-Tone Air Conduction
Masking for air conduction is required whenever the presentation level of the test signal in the test ear minus the interaural attenuation is greater than or equal to the bone-conduction threshold of the non-test ear:
Criteria for Masking Pure-Tone Bone Conduction
Masking for bone conduction is required whenever an air-bone gap (ABG) in the test ear exceeds :
Quantitative Masking Equations and Occlusion Parameters
Effective Masking and Boundary Definitions
Effective Masking (EM): The minimum intensity of a masking noise required to just mask a signal at a given presentation level. EM serves as the starting point for clinical masking.
Overmasking: A condition wherein the masking noise delivered to the non-test ear is so intense that it crosses the skull back to the test ear, raising the threshold of the test ear signal.
Pure-Tone Air Conduction Masking Level Equations
Minimum Masking Level for Air Conduction:
where the standard safety buffer is approximately .
For Insert Earphones ():
For Supra-Aural Headphones ():
Maximum Masking Level for Air Conduction:
For Insert Earphones ():
For Supra-Aural Headphones ():
Pure-Tone Bone Conduction Masking Level Equations
Minimum Masking Level for Bone Conduction:
where the standard safety buffer is approximately .
Occlusion Effect (OE) Values:
The Occlusion Effect is an artificial enhancement of bone-conducted low-frequency sound caused by occluding the external auditory meatus with a transducer (earphone).
Frequency (Hz) | Occlusion Effect (dB) |
|---|---|
Maximum Masking Level for Bone Conduction:
Speech Audiometry Masking Formulas
Masking principles apply similarly to speech signals, using speech noise centered over speech frequencies across the to spectrum.
Speech Recognition Threshold (SRT) Minimum Masking Level:
Word Recognition Score (WRS) Effective Masking Level:
where is the Presentation Level in the test ear.
Maximum Speech Masking Level:
Largest Significant NTE ABG and Best TE BC refer to values measured within the octave range of to .
The Plateau Method for Threshold Determination
The Plateau Method is the standard clinical technique for determining true masked threshold levels in pure-tone audiometry.

Plateau Mechanics and Regions
Undermasking: When masking noise presented to the non-test ear is insufficient, any increase in masking noise causes a corresponding linear increase in the perceived test ear threshold because the non-test ear is still hearing the test tone.
Minimum Effective Masking: The lowest masking noise level that effectively eliminates the non-test ear's contribution.
The Plateau: A range of masking noise intensities across which the test ear threshold remains constant. An increase of masking noise in the non-test ear across a total range of to (applied in or increments) without a shift in test ear response confirms the true threshold.
Maximum Useable Masking: The upper bound of masking noise before overmasking occurs.
Overmasking: Occurs when the masking noise level exceeds the maximum limit, crossing the skull to elevate the test ear's true threshold. Clinically, overmasking causes a continuous back-and-forth increase in both presentation level and masking level without ever establishing a plateau.
Procedural Steps for the Plateau Method
Present the test ear pure tone at the unmasked threshold level while delivering the calculated minimum masking level to the non-test ear continuously.
If the tone is unheard, raise the pure-tone intensity in the test ear in steps until the patient responds.
Once the patient responds to the tone in the presence of masking noise, increase the masking noise level in the non-test ear by .
Re-test the tone at the same level in the test ear.
Repeat steps 3 and 4 until the masking noise is increased by a total of to (e.g., three consecutive steps) while the patient continues to perceive the pure tone at the exact same intensity level in the test ear.
Record the final test tone intensity as the true masked threshold.
Record the final masking intensity level (the level at the end of the plateau, not the initial minimum masking level) in the masking documentation table on the audiogram.
Worked Examples and Clinical Case Studies
Case Study 1: Headphones Pure-Tone Evaluation
Test Parameters: Supra-Aural Headphones (), Test Frequency = .
Audiometric Findings:
Right Ear (TE / AD): Unmasked Air Conduction () = , Unmasked Bone Conduction () = .
Left Ear (NTE / AS): Unmasked Air Conduction () = , Unmasked Bone Conduction () = , Air-Bone Gap () = .
Step 1: Check Need for Masking
Air Conduction Check:
Since , Air Conduction masking is required for the Right Ear.
Bone Conduction Check:
Since , Bone Conduction masking is required for the Right Ear.
Step 2: Calculate Air Conduction Masking Levels (Right Ear)
Minimum Masking Level:
Masking noise starting level in Left Ear = .
Maximum Masking Level:
Plateau range available with headphones = to (a range before risk of overmasking).
Step 3: Calculate Bone Conduction Masking Level (Right Ear at 1000 Hz)
Minimum Masking Level:
Case Study 2: Speech Audiometry Evaluation
Test Parameters: Supra-Aural Headphones ().
Audiometric Findings:
Right Ear (TE / AD): Speech Recognition Threshold () = , Word Recognition Presentation Level () = .
Left Ear (NTE / AS): , Largest Significant .
Step 1: SRT Masking Determination
Check Crossover:
Since , SRT masking is required for the Right Ear.
Step 2: WRS Minimum Masking Level Calculation
Calculation:
Case Study 3: Detailed Plateau Method Walkthrough (Bone Conduction)
Test Parameters: Insert Earphones (), Frequency = .
Audiometric Findings:
Left Ear (TE / AS): Unmasked Air Conduction = , Unmasked Bone Conduction = ().
Right Ear (NTE / AD): Expected Bone Conduction = , .
Step 1: Verify Masking Requirement
Since , Left Ear Bone Conduction masking is required.
Step 2: Minimum Masking Level Calculation
Step 3: Plateau Procedure Execution
Initial Trial: Present BC tone at to Left Ear with continuous masking noise at in Right Ear.
Patient does not hear tone.
Tone Threshold Search: Keep masking noise at . Increase Left Ear BC tone in steps until patient responds.
Patient responds when Left Ear BC reaches .
Plateau Step 1: Increase masking noise by (to ). Present Left Ear BC tone at .
Patient confirms hearing tone at .
Plateau Step 2: Increase masking noise by (to ). Present Left Ear BC tone at .
Patient confirms hearing tone at .
Plateau Step 3: Increase masking noise by (to ). Present Left Ear BC tone at .
Patient confirms hearing tone at .
Step 4: Final Documentation and Overmasking Check
True Masked Threshold: for Left Ear Bone Conduction.
Audiogram Notation: Change unmasked BC symbol (
>) to masked BC symbol (]). Record the end-of-plateau masking level () in the masking table under AS BC.Overmasking Verification:
Since final plateau level () < Maximum Masking Level (), the measurement is verified accurate with zero overmasking danger.