Exam Tidbits: Lead Fractures and Pseudo Fractures
Conductor Fracture Clinical Signs
- Rising pacing impedance.
- Intermittent loss of sensing and/or pacing.
- Complete failure of sensing and/or pacing.
- Oversensing (contact potentials).
- Source: Gradaus, Rainer, et al: ICD Leads: Design and Chronic Dysfunctions; PACE 2003; 26 (Pt. I) 649-657
Lead Fractures
- Most information is derived from "Cardiac Pacing" edited by Kenneth Ellenbogen, specifically the chapter "Differential Diagnosis, Evaluation, and Management of Pacing System Malfunction" by Paul A. Levine, MD.
- Levine appropriately titles the section dealing with lead fractures as OPEN CIRCUITS, as what appears to be a lead fracture may be due to other issues such as:
- Improperly tightened set-screw.
- Device being past its End of Life (EOL), resulting in no output.
- Lead fractures can manifest as constant or intermittent events.
- Even with a complete lead fracture, the ends of the wire may remain close enough for intermittent contact and capture.
Bipolar Wires
- Inner coil (cathode) fracture:
- Results in an open circuit.
- No pacing or sensing occurs.
- Reprogramming to unipolar mode will not resolve the issue.
- Outer coil (anode) fracture:
- Results in an open circuit.
- Reprogramming the device to unipolar mode may provide temporary functionality because unipolar pacing makes the device ignore the anode.
- Short circuit:
- If the inner insulation opens due to stress (e.g., from a suture), a short circuit can occur between the anodal and cathodal wires.
- Results in no pacing or sensing, or intermittent capture and sensing if the defect is intermittent (in unipolar configuration).
Diagnosing Lead Fractures
- Chest X-ray (AP and Lateral):
- Fractured unipolar wires are generally easier to visualize than bipolar wire fractures.
- Fractures may not always be visible on CXR, especially with bipolar leads.
- Device-based electrograms (if available):
- Look for make/break signals, which indicate intermittent contact.
- Look for total signal drop-out.
- Caution with Provocative Testing: Be very careful with provocative testing using arm and/or shoulder motion, as an intermittent fracture could become a complete fracture, potentially leaving a device-dependent patient with no rhythm.
Usual Sites of Transvenous Lead Fracture
- Site of venous entry: 40%
- Between the site of venous entry and the generator: 28%
- Close to the lead insertion into the generator: 23%
- Intravascular: 7%
- Source: Alt E, Volker R, Blomer H. Lead fracture in pacemaker patients. Thorac Cardiovasc Surg 1987; 35:101-104.
- Lateral and Posteroanterior chest X rays 3 months after implantation showing complete transection of the ventricular lead. The proximal end of the distal segment migrated to the right pulmonary artery (arrow).
Pseudo Fracture
- Pseudo fractures are not limited to ICD leads; they can also be observed in some bifurcated bipolar pacing leads.
- Pseudo fractures are apparent discontinuities of the coil commonly seen on chest x-ray.
- The images presented show examples of normal leads that may appear damaged.
- Left image source: Camos, Jean Pierre, et al; PACE 1999; 22:1704. Pseudofracture of an ICD Lead.
- Right image source: Kratz, John M. et al; PACE 1995; 18:2225-2226. Pseudo Fracture of a Defibrillating Lead.