Internal Quality Control (IQC) – Comprehensive Study Notes
Internal Quality Control (IQC): Definition, Purpose & Scope
- WHO (1981) definition: “A set of procedures for continuously assessing laboratory work and emergent results.”
- Operates under laboratory-specific IQC protocols while meeting NATA obligations.
- Core objective: guarantee day-to-day consistency so patient results are reliable before release.
- Patient reports are issued only when IQC fulfils pre-established acceptance criteria.
- Two complementary pillars of analytical-phase validation:
- Internal Quality Control (current focus)
- External Quality Assurance (EQA / proficiency testing) – covered in next lecture.
Control Materials (QC Materials)
- Commercial examples: Bio-Rad frozen, freeze-dried (lyophilised) or chemically-preserved vials with manufacturer-assigned target values.
- Lyophilised vials require reconstitution:
- Strict adherence to protocol; accurate pipetting of diluent critical to restore correct analyte concentration.
- Non-routine analytes may use in-house pooled patient sera:
- Laboratory establishes its own target values via senior-scientist verification studies.
- Success of IQC hinges on selecting appropriate material.
ISO 15189 Requirements for Selecting QC Material
- Concentration range must cover clinical decision points encountered in patient samples.
- Minimum two levels (e.g. low vs high, normal vs abnormal, positive vs negative, reactive vs non-reactive).
- Matrix should mimic patient sample composition as closely as possible.
- QC must be analysed identically to patient samples; errors in the run must manifest in QC data.
- Example pitfall: never run a urine QC on a serum assay.
- Minimise variability:
- Long-term stability, minimal vial-to-vial / batch-to-batch differences.
- Hence commercial, certified QCs preferred over in-house preparations.
Establishing Control Limits
- Laboratories usually derive limits from their own data.
- Data collection guideline:
- At least 20 independent data points from 20 separate, stable operating days.
- Include all results—even failed values—to capture true precision.
- Statistical groundwork:
- Calculate mean (\overline{x}) and standard deviation (SD).
- Assume Gaussian distribution.
- Conventional control limits: \overline{x} \pm 2SD ⇒ 95\% confidence; 5\% probability a value randomly falls outside.
Levy-Jennings (LJ) Charts
- QC result plotted chronologically; vertical axis = result, horizontal axis = time/run number.
- Horizontal lines mark \overline{x},\; \pm1SD,\; \pm2SD,\; \pm3SD.
- Two charts usually maintained in parallel:
- Level 1 (control 1) – typically low/normal range.
- Level 2 (control 2) – high/abnormal range.
- Scientists interpret patterns weekly/fortnightly/monthly for long-term precision, accuracy and bias trends.
Westgard Rules (Two-Level QC System)
Decision matrix for accepting or rejecting analytical runs.
Numerical prefix = number of results; “S” subscript = SD distance.
Rule | Meaning | Action | Error Type Clue |
---|
1-2S | 1 result > \pm2SD | Warning only | — |
1-3S | 1 result > \pm3SD | Reject run | Random |
2-2S | 2 consecutive results > \pm2SD on same side (within or across levels/runs) | Reject | Systematic |
R-4S | Range of 4SD within same run (one +2SD, one –2SD) | Reject | Random |
4-1S | 4 consecutive results > \pm1SD on same side (within/across) | Reject | Systematic |
10X | 10 consecutive results on same side of mean (within/across) | Reject | Systematic |
Always evaluate both levels (within-run & across-run) before releasing patient data.
Interpreting QC Data: Precision, Accuracy & Bias
- Precision = reproducibility (tight clustering).
- Accuracy = closeness to true/target value.
- Bias = consistent deviation (high or low) from target.
- Ideal IQC outcome: precise + accurate (tight cluster at target).
- Scenarios:
- Precise but inaccurate ⇒ bias present (shift on LJ chart).
- Inaccurate & imprecise ⇒ random scatter; invalid assay.
Error Categorisation
- Random Error (Imprecision): sporadic, typically single excursions >3SD.
- Causes: bubbles in QC vial, pipetting variability (worn syringes), electrical interference.
- Westgard triggers: 1-3S, R-4S.
- Systematic Error (Bias): persistent drift or shift from mean.
- Causes: calibration bias, reagent lot changes, instrument deterioration, non-specific assay reaction.
- Triggers: 2-2S, 4-1S, 10X.
Pattern Recognition on LJ Charts
- Trend: ≥5 consecutive points moving steadily up or down.
- Often gradual light-source decay or ageing QC/calibrator material.
- Shift: ≥6 consecutive points abruptly above or below mean.
- Often reagent lot change, faulty recalibration, major maintenance.
Troubleshooting Failed QC
- Do not rerun blindly; perform systematic investigation:
- Inspect QC material: expiry, storage, vial-to-vial variation.
- Check reagent pack: lot changes, contamination, expiration.
- Verify calibrator: value assignment, preparation errors.
- Examine analyser: mechanical faults, optics, probes.
- Review environment: temperature, humidity, electrical stability.
- After identifying cause:
- Rectify (e.g. replace reagent, recalibrate, service analyser).
- Rerun QC to confirm values fall within \pm2SD before releasing patient results.
- Example: A faulty analyser will consistently fail reruns until the hardware issue is fixed.
Limitations of IQC
- Compares laboratory against itself; detects changes but not pre-existing systematic errors embedded in original method validation.
- Slow drift affecting all results may remain invisible if it started before baseline was established.
- Mitigation: participate in External Quality Assurance (EQA) / Proficiency Testing to benchmark against peer laboratories.
Key Takeaways
- IQC is mandatory, continuous and immediate; EQA provides external, periodic benchmarking.
- Proper QC material selection, statistical limit setting, vigilant LJ monitoring and disciplined application of Westgard rules are essential for reliable laboratory reporting.
- Ethical imperative: Never release patient results until QC integrity is unequivocally demonstrated.