Midterm 1-6(14)
summary with headers with brief description
PP1/Unit1
The lecture introduces the main parts, surface patterns, and correct handling techniques for common surgical instruments.
Instrument Parts
Jaws are the working ends that open and close to grasp, cut, or crush tissue and vessels. They may be straight, curved, or tapered.
Teeth are present on some instruments to improve tissue grip and provide a secure closure.
Box lock is the hinge where the two shanks meet. It enables jaw movement, absorbs substantial stress, and should be checked regularly.
Shaft/shank forms the main body and determines overall instrument length, typically about 3–12 inches.
Ratchet is the interlocking locking mechanism that holds the jaws closed during a procedure.
Finger rings are the holes used to hold ring-handled instruments, usually with the thumb and ring finger.
Jaw Serrations
Serrations are grooves in the jaws of instruments designed to hold or crush tissue.
Longitudinal (vertical): grooves run along the length of the jaw.
Transverse (horizontal): grooves run across the jaw.
Cross-serrated: intersecting grooves create a checkerboard pattern.
Proper Instrument Holds
Hemostats and scissors: Place the thumb and third finger in the rings. The middle finger supports the instrument, the index finger guides it, and the pinky adds stability.
Thumb forceps: Hold between the thumb and middle finger, using the index finger for control. The non-dominant hand stabilizes tissue.
Scalpel handle: Use a pencil grip, holding it between the thumb and index finger in the non-dominant hand and resting it on the other fingers.
Core Learning Goal
Correct identification and handling of instrument features supports controlled movement, secure tissue management, and safe surgical technique.
Waste Anesthetic Gas (WAG)
WAG includes vapors from inhalant anesthetics such as nitrous oxide, isoflurane, and sevoflurane that escape into clinic air. Every staff member can be affected, and exposure risk increases with dose and duration.
Health Effects
Short-term exposure can cause drowsiness, headache, fatigue, nausea, itching, depression, and irritability. Repeated symptoms during work can indicate excessive gas levels.
Long-term exposure has been linked to reproductive problems, liver and kidney disease, bone-marrow abnormalities, and possible memory, motor, and neurologic changes. Nitrous oxide has particular reproductive and neurologic concerns.
Main Sources of Exposure
Exposure occurs near operating anesthetic machines and recovering animals, from liquid-agent spills, and from gas released by equipment such as hoses, bags, and reservoirs. Surgery, dentistry, and recovery areas are highest risk.
Factors That Increase Risk
Long procedures, high carrier-gas flow, ineffective scavenging, mask or chamber induction, and poor ventilation raise room concentrations. The document recommends 15–20 air changes per hour.
Exposure Limits
Recommended limits are 2 ppm for halogenated anesthetics, including isoflurane and sevoflurane, and 25 ppm for nitrous oxide. Odor is not a reliable warning: halothane can be smelled only around 33 ppm, far above the recommended maximum.
Core Prevention Measures
Use an effective scavenging system, perform leak tests before use, avoid mask induction and induction chambers, maintain equipment, and use appropriate PPE for unavoidable exposure such as spills. Active scavenging systems use a pump or fan to exhaust gas outdoors and can substantially reduce WAG.
Equipment Testing and Maintenance
Perform a low-pressure leak test at the beginning of each day, when systems change, or after another user operates the machine. Check seals, washers, hoses, bags, valve covers, the CO2 canister, and fittings. Service machines annually or twice yearly, and test ET-tube cuffs with every use.
Safer Anesthetic Technique
Intubate promptly if mask induction is unavoidable, use cuffed ET tubes, use the lowest effective oxygen flow, and do not start the vaporizer until the tube is cuffed and connected. Turn off oxygen and anesthetic before disconnecting the patient, and allow several minutes of oxygen after vaporizer shutoff to clear exhaled gas through the system.
Spill and Respiratory Protection
For a spill, evacuate personnel, maximize ventilation, wear gloves, apron, and an organic-vapor respirator, then absorb and dispose of the material in an airtight container outside the clinic. N95 masks provide no protection against WAG; only fit-tested organic-vapor respirators are appropriate.
Oxygen Cylinder and Vaporizer Safety
Secure oxygen cylinders upright with chains or belts, keep valve caps on unused tanks, use a handcart for transport, and open connected valves slowly while facing away from the outlet and gauge. Keep vaporizers upright; if tipped, allow 12–24 hours for agent to settle and consider recalibration before use.
CO2 Absorbent Safety
Replace soda lime monthly or after 6–8 hours of use, even if its color has not changed. Hard or dry granules, temporary purple color changes, and increased inspired CO2 or ETCO2 can indicate exhaustion.
Injectable Anesthetic Safety
Injectable agents can be harmful through skin contact, eye splashes, or ingestion. Potent opioids can cause respiratory failure and death; safe practice requires training, never working alone, PPE, reversal agents, first-aid supplies, an emergency plan, and immediate sharps disposal.
PP2 ~ Unit1
This unit explains how to identify, select, and safely handle common cutting instruments and thumb forceps used in veterinary surgery.
Scalpel handles and blades
Scalpels make precise tissue cuts with minimal injury to nearby structures. No. 3 handles are used for small-animal blades numbered below 20, typically beginning with “1,” while No. 4 handles fit large-animal blades numbered 20 or higher, typically beginning with “2.” Common veterinary blades include Nos. 10, 11, 15, and 20.
Safe scalpel blade handling
Attach and remove blades with a needle holder or forceps rather than fingers. Keep the cutting edge pointed away from yourself and others, ensure the blade is seated flush on the handle, and direct the blade downward toward the trolley during removal.
Scissors: purpose and classification
Scissors cut tissue, perform blunt dissection, and cut sutures. They must remain sharp to avoid crushing tissue. They are classified by purpose/name, blade shape (straight or curved), and tip type (blunt/blunt, sharp/sharp, or sharp/blunt).
Main scissor types
Operating scissors: Cut inanimate materials such as gauze and sutures; identified by blade shape and tip type.
Mayo scissors: Thick, blunt/blunt scissors for large muscle masses, cartilage, and other non-delicate tissue.
Metzenbaum scissors: Long, fine instruments with short thin blades for delicate dissection.
Suture scissors: Blunt/blunt scissors for cutting sutures; Littauer scissors have a hook to lift the suture before cutting.
Bandage scissors: Have a flat, blunt lower jaw that protects the skin while cutting bandage material.
Iris scissors: Small, fine scissors designed for eye work; available with sharp/sharp or blunt/blunt tips.
Thumb forceps
Thumb forceps pick up and hold tissue or dressing materials. Their spring action requires the user to close the jaws manually, and they should be held like a pencil between the thumb, index, and middle fingers.
Tissue versus dressing forceps
Tissue forceps have teeth, allowing a secure short-term hold on tissue with less crushing and slippage. Examples include rat-tooth, Adson, and Adson Brown forceps. Dressing forceps have blunt, serrated, toothless tips for applying or removing dressings and handling hollow organs without damage.
PP3 ~ Unit1
This PowerPoint explains the design, identifying features, and veterinary surgical uses of hemostatic forceps, tissue-grasping forceps, needle holders, and towel clamps. It emphasizes selecting the smallest and least traumatic instrument appropriate for the tissue or task to minimize surgical damage.
Hemostatic Forceps
Hemostatic forceps clamp blood vessels to control bleeding. They have ratchets and box locks, come in straight or curved forms, and should be used in the smallest effective size while gripping only the tissue required.
Halstead mosquito: Small, delicate forceps with full-length fine transverse serrations; used for small vessels.
Kelly: About 5 inches long, with serrations on the distal two-thirds; used for medium vessels or small tissue masses.
Crile: Similar to Kelly forceps, but serrations run the full jaw length.
Kocher: Full-length transverse serrations plus 1 × 2 teeth; firmly grasps medium-to-heavy tissue and dense vessels.
Rochester Pean: Large, toothless forceps for large vessels and muscle or tissue masses.
Rochester Ochsner: Like Rochester Pean but with teeth; also resembles a larger Kocher.
Rochester Carmalt: Large forceps with longitudinal jaw serrations and cross-serrated tips; clamps large vessels or tissue masses with less crushing than transverse serrations.
Ferguson angiotribe: A highly traumatic crushing instrument for nonviable tough tissue bundles, such as a uterine stump during a spay.
Tissue-Grasping Forceps
These ratcheted instruments hold tissue during surgery. Although designed to limit damage, they are still traumatic and should be chosen according to tissue fragility.
Allis: Short opposing teeth stabilize tough tissue such as fascia or connective tissue. Avoid use on intestines because it may cause damage.
Babcock intestinal: Curved looped jaws with fine longitudinal striations; used to hold intestines and muscle groups with less trauma than Allis.
Doyen intestinal: Fine serrations and jaws that do not fully close; used on intestinal loops or to prevent intestinal spillage with minimal trauma.
Foerster sponge: Holds sponges or gauze for surgical preparation and blood absorption during surgery.
Needle Holders/Drivers
Needle drivers are designed to hold curved suture needles while suturing or tying knots. Cross-hatched jaw tips reduce needle slippage, and some models include suture-cutting scissors.
Olsen-Hegar: Combines a needle holder and scissors, reducing instrument changes but creating a risk of accidentally cutting suture material.
Mayo-Hegar: Needle holder only; a grooved, serrated tip improves needle grip.
Mathieu: Needle holder only, usable with either hand; its ratchet releases when added pressure is applied to the spring handles.
Towel Clamps
Towel clamps secure towels and surgical drapes to each other and, in some designs, to the patient. Most have pointed ends intended to pierce skin.
Backhaus: The most common type; needle-like tips and rounded jaws provide firm attachment.
Roeder: Steel balls on the jaws limit deep tissue penetration and help prevent towel slippage.
Edna/Lorna: Wide jaws with fine non-perforating teeth; secure drapes, towels, or tubing without penetrating the drape or skin.
Core Principle
Instrument selection should match tissue size and fragility: use the least traumatic instrument and the smallest effective size that accomplishes the surgical task.
PP4 ~ Unit1
The document covers surgical retractors, delicate ophthalmic instruments, electrosurgery safety, and surgical suction.
Retractors
Retractors hold incision edges apart to improve access and visibility. They may be hand-held, requiring a scrubbed assistant, or self-retaining, using locking mechanisms such as ratchets or screws. Moist sponges or towels can protect tissues when abdominal or thoracic retractors are used.
Hand-held: Snook spay hook retrieves uterine horns during an ovariohysterectomy; Senn and Volkman retractors have blunt or toothed/pronged ends; Hohmann retractors are commonly used in orthopedics.
Self-retaining: Weitlaner and Gelpi retract muscles or body cavities; the Gelpi’s pointed tips can cause tissue trauma. The Balfour retractor keeps abdominal walls open for wide abdominal exposure.
Ophthalmic instruments
Eye-surgery instruments are small, fine, delicate, and comparatively expensive, so they must be handled carefully and kept separate from standard instruments.
Thumb forceps and iris scissors: Used for delicate tissue handling and cutting.
Fine needle drivers: Such as Castroviejo drivers, designed for precise suturing.
Eye speculum: Holds the eyelids apart.
Hemostasis supplies: Small gauze, sterile cotton-tipped applicators, and eye sponges control or remove small amounts of blood.
Electrocautery and electrosurgery
Electrocautery heats a wire electrode to coagulate small vessels or destroy tissue; the current does not pass through the patient. Electrosurgery sends high-frequency alternating current through patient tissue to cut, coagulate, fulgurate, dissect, or ablate.
Benefits: Less bleeding, reduced postoperative swelling, shorter procedures, and less anesthesia time.
Risks: Thermal injury, staff shock, surgical fire, and hazardous surgical smoke.
Monopolar units: The most common type; a handpiece and grounding plate complete the circuit and allow cutting or coagulation.
Bipolar units: Current passes only between two tips; they coagulate but do not cut, making them useful for small vessels near sensitive tissues.
Electrosurgery safety
Safe operation depends on correct grounding and minimizing alternate electrical pathways. The grounding plate should have 85–95% contact with the patient, with conductive gel applied. Avoid skin-to-skin contact, remove metal collars or tags, and do not use flammable alcohol-based prep agents. Staff should not touch the patient during activation, and smoke evacuation should be used close to the plume because surgical smoke is hazardous.
Suction
Suction removes fluid or air to improve visibility, clear blood, and remove lavage fluid from incisions or joints. It may be manual or mechanical.
Manual suction: Includes syringes, bulb syringes, and Jackson-Pratt drains, which create suction when the bulb is compressed.
Mechanical suction: A vacuum pump draws fluid through tubing into a collection container; excessive vacuum can damage tissue.
Common tips: Poole for abdominal and thoracic procedures, Frazier for orthopedic procedures, and Yankauer for general-purpose suction.
Technologist responsibilities
The technologist may hold a hand-held retractor or suction tip while scrubbed in, operate suction controls while non-sterile, set up and take down equipment, ensure safe electrosurgical preparation, and dispose of collected fluid properly
PP1/Unit2
Proper cleaning, inspection, and handling preserve instrument function and help prevent staining, pitting, corrosion, and surgical delays.
Instrument Quality and Finish
Quality affects durability and cost. Higher-priced instruments generally use higher-quality steel, while lower-quality instruments rust, corrode, pit, and break sooner. Processing low- and high-quality instruments together can damage the higher-quality items.
Finish affects care needs. Satin instruments reduce operating-room light glare but are more prone to spotting and discoloration than shiny finishes, so they require especially diligent cleaning.
Corrosion Prevention
Residue, tap water, and unsuitable cleansers are the main risks. Blood, pus, and other secretions contain chloride ions that can stain or corrode instruments, particularly when left for 1–4 hours or baked on during autoclaving. Mineral-heavy tap water can leave deposits and permanent stains; distilled water should be used after manual washing. Dish soap, laundry soap, and surgical hand scrub can cause spotting and corrosion.
Clean promptly. Begin washing within 10 minutes after surgery, remove visible debris immediately, and never sterilize an instrument until it is clean.
Rust Versus Stain
Use the eraser test to identify discoloration. If a pencil eraser removes a brown or orange mark and the metal beneath is smooth, it is a stain. If a pit remains, it indicates corrosion or rust and permanent damage.
Inspection Standards
Confirm cleanliness, condition, and function before reuse. Check forceps-tip alignment; ensure hemostats and needle holders close without light between their jaws and lock/unlock smoothly; inspect needle-holder jaws for wear; and verify that cutting edges are sharp, undamaged, and that scissors glide smoothly without loose closed blades.
Cleaning Process
Use approved enzymatic and ultrasonic cleaning methods. Enzymatic cleaners contain surfactants and enzymes that break down blood, tissue, proteins, lipids, and bone residue. The standard sequence is:
Rinse off large debris and blood immediately after use.
Soak briefly in a commercial instrument cleaner.
Gently scrub with a soft brush under running water.
Ultrasonically clean for 5–10 minutes with the appropriate solution, following manufacturer instructions.
Rinse, immerse in instrument lubricant, and air-dry.
Use the ultrasonic cleaner correctly. Remove visible debris first, avoid overloading, and open ratchets and box locks so all surfaces are exposed. Ultrasonic cleaning is described as substantially more effective than hand cleaning.
Lubrication and Cold Sterilization
PP1 ~ Unit3
Sterilization prevents surgical contamination by eliminating microorganisms, including bacterial spores, through correctly prepared packs and controlled autoclave cycles.
Core infection-control terms
Pathogens are microorganisms capable of causing disease, including viruses, bacteria, parasites, fungi, and yeasts. Infection occurs when pathogens establish and multiply, such as after contaminating a surgical site.
Disinfectants are used on non-living surfaces and reduce or destroy many viruses, fungi, and bacteria, but not spores. Antiseptics are used on living tissue and destroy bacteria and some viruses, but not spores. Aseptic technique consists of precautions that prevent contamination.
Sterilization methods
Four methods are described:
Filtration: Traps microorganisms in liquids or gases; used for pharmaceuticals.
Radiation: Uses energy to destroy microorganisms without high temperatures; suitable for heat-sensitive items such as sutures and gloves.
Gas sterilization: Uses toxic ethylene oxide gas for heat-sensitive materials such as plastic and rubber; requires specialized equipment.
Thermal heat: Destroys microorganisms by protein denaturation. Steam sterilization is the most common form.
How an autoclave sterilizes
An autoclave heats distilled water to create steam. The steam pushes air out of the chamber, pressure rises until the operating temperature is reached, and steam penetrates the packs. After the cycle, the steam condenses and is removed.
Effective sterilization generally requires 121–134°C (250–274°F), 15–30 psi, and 7–20 minutes of cycle time, excluding drying time. The exact setting depends on the autoclave model and load.
Proper loading and drying
Packs must be clean, correctly wrapped, and placed vertically with 1–3 inches of space between packs and from chamber walls. Avoid overloading so steam can circulate and penetrate every pack.
After the cycle, leave the door cracked open for 10–15 minutes to release moisture. A damp pack is not considered sterile because contaminants can wick through the material.
Pack-preparation requirements
Autoclaves sterilize but do not clean, so instruments, linens, wraps, and drapes must be clean and free of grease or oil before loading. Packs should be no larger than 30 × 30 × 50 cm and weigh no more than 5.4 kg.
Each pack needs an adhesive seal and a label listing its contents, date, and initials. Sterilize most ratcheted instruments in the open position to improve steam penetration; towel clamps and scissors are exceptions.
PP2 ~ Unit3
Sterile packs remain safe only when they are correctly wrapped, properly processed, protected during storage, and opened without contaminating the field.
Pack-wrap requirements
Purpose: A wrap must let steam or gas enter during sterilization while blocking microbes and dust afterward.
Selective permeability: Allows sterilant penetration but prevents outside contamination.
Resistance: Resists handling damage; tears, holes, and worn areas should be visible.
Flexibility: Conforms around the pack’s shape.
Memory: Returns to a flat position, helping prevent accidental contamination when opened.
Wrap types
Textile wraps such as cotton or muslin are flexible and more resistant to tears and punctures, but require laundering and can produce lint. Damaged textile wraps must be discarded.
Paper/plastic peel pouches are convenient for individual or small items and need no laundering. They are single-use, relatively expensive, and may dry out or crack.
Sterilization indicators
Indicators verify that processing conditions were achieved; they do not replace proper wrapping and handling.
External chemical tape: Seals the pack and shows that steam/temperature was present. It is also used to record pack contents, date, and wrapper initials.
Internal chemical strip: Placed beside instruments or inside soft goods; it changes color when sterilization conditions, including temperature and pressure, are reached.
Biological indicators: Provide the strongest verification by showing whether spores or bacteria were destroyed under the required temperature, pressure, and time. Spore strips are more common; vial tests require incubation.
Place a chemical or biological indicator in the middle of the pack and seal the exterior with steam-indicator tape.
Single versus double wrapping
Double wrapping provides an added contamination barrier. A non-sterile person can remove the outside layer while a sterile person opens the inner layer, helping preserve sterility. With gowns and towels, both layers may be opened.
Wrapping method
Clean items first, then use the angle-fold method for cotton or disposable wraps.
Lint-roll cloth materials and fold towels, drapes, and gowns with an accordion pleat so they unfold safely.
Center the item on a clean wrap.
Bring up the bottom flap and fold back a corner to form a tab.
Fold each side flap to the center, creating tabs.
Make the final fold, tuck it under the other flaps, and leave a short tab for easy opening.
Place an internal indicator strip and secure the pack with autoclave tape.
Storage and shelf life
Store dated packs where they remain clean, dry, and protected from damage.
Single-layer cotton or non-woven packs: 1 week in a closed cupboard; 2 days in an open cupboard.
Double-layer cotton packs: 7 weeks closed; 3 weeks open.
Peel-away pouches: at least 1 year.
Sterility is also event-related: a pack is no longer considered sterile if it becomes wet, torn, punctured, dropped, or dusty. Before use, check both its date and physical condition; if either is questionable, re-sterilize it.
Opening sterilized packs
Maintain an 18-inch separation from the sterile field and prevent contact with the contents. Keep hands and arms above waist level, do not reach across the pack, and handle wrappers only by their corners. A non-sterile person generally opens the outer layer of instrument and drape packs, while a scrubbed person opens the inner layer.
PP1 ~ Unit4
This unit introduces 11 orthopedic instrument categories and explains internal versus external fracture fixation.
Learning Objectives
Students should be able to identify orthopedic instruments, distinguish internal from external fixators, describe advantages of intramedullary (IM) pins, and recognize when external fixation is needed and its complications.
Bone Handling and Preparation Instruments
Bone holders: Clamps that hold bone fragments together during surgery.
Periosteal elevators: Lift periosteum away from bone to allow implant placement; also used for periosteal stripping in foals with angular limb deformities.
Bone rongeurs: Sharp, cupped instruments that remove or break up bone pieces for grafting or removal.
Bone rasps/burrs: Used to smooth or shape bone.
Bone curettes: Remove small pieces of bone.
Bone chisels/osteotomes: Cut or chisel bone.
Intramedullary Pinning
Jacob’s hand chuck: Holds and drives IM pins into bone; a key tightens the pin, and an extension protects the surgeon’s hand.
Steinmann IM pins: Stainless-steel pins, 1/16 to 1/2 inch in diameter, with smooth and/or threaded ends. They may be inserted using a hand chuck or electric drill.
IM pins provide internal fixation by reducing and realigning fractures. They nearly fill the intramedullary canal and temporarily affect blood-cell supply within it, which regenerates within about one week. Two pins may be used as stacked pins.
Advantages of IM Pins
IM pinning offers stability for simple fractures, low cost, minimal equipment requirements, and a relatively easy technique to learn.
Orthopedic Wire, Plates, and Screws
Cerclage wire: Stainless-steel wire intended for long-term implantation, used with plates, pins, or screws; it can also repair some jaw fractures alone.
Bone plates and screws: Stainless-steel implants used for internal fixation of fractures. Plates come in varying sizes and shapes, including straight, T-, and L-plates. Plate width corresponds to the screw size needed for fixation.
External Fixation
External fixators are mainly used for complicated fractures and fractures associated with open wounds. The most common type is the Kirschner apparatus, which uses pins extending into bone and a bridging bar that stabilizes the pins; the pins may be bandaged.
Complications of External Fixators
Potential problems include pin loosening, pin-site infection, bent pins, and pin movement. Other external stabilization methods include casting and splinting.
Key Distinction
Internal fixation places implants such as pins, plates, screws, or wire within the body, while external fixation stabilizes the bone through pins connected to a structure outside the body.
PP1 ~ Unit5
Proper surgical preparation protects the patient from infection by controlling contamination, preparing the environment, and using correct clipping and patient-prep practices.
Purpose and learning goals
This unit explains how to prepare surgical patients, personnel, and surgical spaces while maintaining aseptic technique. It covers microorganism sources, surgical conscience, room setup, clipping, surgical-site preparation, and bladder emptying.
Aseptic technique and surgical conscience
Asepsis makes the surgical environment surgically clean, though not sterile, and is essential for infection prevention.
Endogenous microorganisms originate within the patient, such as from the gums or intestines, and cannot be fully controlled.
Exogenous microorganisms come from outside the patient, including instruments, personnel, and skin; these are largely controllable through proper technique.
Surgical conscience means every team member consistently follows aseptic procedures and immediately reports and corrects any break, even when no one else observes it.
Personnel preparation
The surgical team must limit hair and airborne contamination before patient preparation begins.
Wear a lab coat or scrub jacket over scrubs while clipping and vacuuming, then remove it afterward.
Wear only clean scrubs for surgery.
Put on surgical caps and masks when the surgical prep begins to prevent hair and respiratory contamination.
Preparation and surgery areas
The prep room and surgery room have different functions but both must support cleanliness and controlled traffic.
Preparation area: Ideally adjacent to surgery; used for clipping, patient preparation, supply storage, and possibly surgeon scrubbing and gowning. Dirty procedures, such as wound care or abscess treatment, belong here rather than in the surgery room.
Surgery room: Reserved for aseptic procedures. Doors should close to reduce traffic, and surfaces and materials should be nonporous and easy to clean.
Room setup: Confirm the table is clean, use a clean towel and warming device as appropriate, set up instrument tables and packs, check lights, start monitors, and prepare the anesthetic machine.
Equipment control: Keep equipment in the surgical suite whenever possible. If movable equipment leaves, clean and disinfect it before returning.
Patient verification and surgical-site preparation
Before preparation, verify the correct patient, procedure, and sex. The RVT is identified as the primary person responsible for patient prep.
Surgical preparation includes removing hair, cleaning loose hair promptly, and creating a sufficiently large, neat, symmetrical clipping field. A general guideline is to clip about 4 inches in every direction from the planned incision and to clip more rather than less.
Common clipping areas and incisions
Feline castration: Do not shave; pluck hair from directly over the scrotum.
Canine castration: Clip cranial to the testicles and about three-quarters of the way up the prepuce, extending laterally at least two clipper widths from midline. Do not clip directly over the testicles because this sensitive area is prone to razor burn.
Canine/feline abdominal procedures: Clip from the xiphoid process to the pelvic brim and flank to flank. The ventral midline incision provides broad abdominal access through the linea alba.
Paramedian incision: Used when access to one side of the abdomen is needed, such as for a retained testicle; it generally requires multilayer closure.
Median sternotomy: A ventral thoracic midline incision used for many heart and lung procedures.
Correct clipper technique
Proper technique reduces skin injury and improves the surgical field.
Hold clippers with a pencil-like grip and keep the blade flat against the skin.
Keep skin taut, especially over curved or contoured areas.
For dense coats, begin with a #10 blade moving with the hair growth to remove most hair efficiently.
Switch to a #40 surgical blade and clip against hair growth for a close surgical clip. If a #10 is not used, make one pass with the grain and a second pass against it.
Remove clipped hair as it is produced, vacuum immediately afterward, use a lint roller for remaining hair, and clean, lubricate, and regularly sharpen clippers.
Razor burn risks
Razor burn can cause irritation, postoperative inflammation, increased licking, and a higher possibility of infection. Flat blade contact, taut skin, and careful handling around sensitive or contoured areas help prevent it.
Bladder check and expression
C
heck and empty the bladder after clipping and hair cleanup but before scrubbing the patient.
Emptying the bladder creates more abdominal working space, reduces the risk of accidental puncture, and helps prevent urine contamination if the relaxed bladder sphincter releases under anesthesia. Express gently by placing the thumb and fingers on opposite sides of the bladder and squeezing in the direction of urine flow; do not apply excessive pressure, and stop if urine does not flow easily.
Overall, consistent aseptic technique, controlled room preparation, precise clipping, and safe bladder management reduce preventable surgical contamination and complications.
Lubricate immediately after cleaning. Instrument lubricant helps prevent rust and corrosion while keeping moving parts functional. Immerse instruments, do not wash after lubricating, and allow them to air-dry.
Cold sterilization has limited use. It means soaking instruments in a disinfectant, such as glutaraldehyde, and may be acceptable for basic disinfection in about 10 minutes. It is not recommended for high-quality surgical instruments because prolonged soaking can cause damage; instruments should never remain in a solution longer than 20 minutes. Glutaraldehyde is toxic and can irritate the eyes and respiratory system.
PP2 ~ Unit5
The document outlines how to safely prepare an animal’s surgical site, transfer and position the patient, and apply a final sterile prep.
Preparation before scrubbing
Remove lab coats and put on a cap and mask before beginning. The surgical-site preparation has three stages: an initial antiseptic scrub, a rinse or second antiseptic, and a final prep.
Initial surgical scrub
Use either chlorhexidine soap or povidone-iodine surgical scrub. Chlorhexidine is non-irritating with rapid antimicrobial action; povidone-iodine is more drying, may stain, and is commonly known as Betadine.
Apply scrub soap with gauze, keeping the bottle tip from touching the gauze. If the animal is dirty, clean the skin until the gauze comes away clean before considering the surgical scrub started. Create a lather and maintain skin contact for at least 2 minutes, using gentle pressure to avoid increased bleeding.
Scrubbing technique and patterns
Start at the planned incision, the cleanest area, and move outward toward the hairline with straight strokes. Do not return from a dirty area to a cleaned area or cross arms over the prepared field. Repeat the scrub three times, or longer if gauze still shows dirt.
Target pattern: Circular, bull’s-eye motion from incision outward; used for abdominal and thoracic procedures.
Orthopedic pattern: With the limb suspended, scrub circumferentially from the incision site down the limb; rinse the same way.
Perineal pattern: Perform separate target patterns on the right side of the anus, left side, and anus itself. A purse-string suture is placed before prep and removed after surgery; do not apply alcohol directly to the anus.
Rinsing or second antiseptic
Use alcohol or sterile water/saline. Alcohol provides a second disinfectant barrier and removes soap residue, while sterile water or saline only rinses away soap.
Apply alcohol with gauze without holding the bottle or arm over the prepped site. Do not soak the patient because alcohol dries and irritates skin and can lower body temperature. Leave alcohol on the area for at least 2 minutes.
Transfer, positioning, and restraint
Before final prep, support the animal’s neck with one hand and pelvis with the other to prevent the hind legs from closing together. If the legs touch, the site must be re-prepped. A sterile huck towel can help protect the prepared area during transfer.
Place the patient carefully on a heated V-trough table. Common positions are dorsal recumbency (on the back), lateral recumbency (on the side), and ventral recumbency (on the chest). Securement practices vary: clinics may tie selected patients, tie all patients, or use sandbags for stability.
When tying, secure front limbs above the elbow and carpus, and hind limbs above and below the tarsus. Attach ropes to adjustable table clamps using a figure-8. Avoid tying the front legs tightly against the chest or above the head because this restricts deep breathing.
Final preparation
The final prep creates a third disinfectant barrier using an alcohol-based iodophor (such as Betadine or Prepodyne), a chlorhexidine-based product, or alcohol.
Apply the solution with sterile sponge forceps or sterile-gloved hands and gauze. Starting at the midline/incision site, use circular motions outward and cover the entire clipped area. Keep gauze away from the hairline to prevent dragging contaminants into the surgical field. Allow the solution to remain for about 2 minutes. After final prep, only scrubbed-in personnel should touch the area before draping.
PP3 ~ Unit5
The document explains how to establish and maintain a sterile surgical field, select and apply drapes, and prepare patients for orthopedic, back, and large-animal procedures.
Purpose of surgical draping
Draping creates a sterile physical barrier that prevents microorganisms from non-prepped skin and the operating table from entering the surgical field. A scrubbed-in team member applies the drapes, leaving only the incision area exposed. The barrier must stay intact throughout surgery; wet or displaced drapes may need replacement.
Drape materials and key risk
An appropriate drape is strong, flexible, comfortable, water-resistant, and easy to sterilize.
Disposable paper: Water-resistant and can be cut to size, but costly.
Plastic adhesive: Pre-sterilized, conforms well to limbs, and is common in orthopedic and exotic procedures.
Muslin/cotton: Economical and often fenestrated, but should not be used where it may become wet.
Strike-through occurs when a cloth drape becomes wet, allowing bacteria to wick through the material and contaminate its surface.
Draping techniques
Two standard methods are used:
Fenestrated drape: A sterile drape with an opening is placed over the site and secured using towel clamps.
Four-quarter/corner drape: Four drapes are placed sequentially clockwise or counterclockwise and secured with towel clamps. Towel drapes may be placed beneath a fenestrated drape to reduce wetness.
Drapes should cover the patient and exposed table surfaces adequately, while the fenestration must be large enough for the planned procedure.
Specialized surgical preparation
Orthopedic limb surgery: Clip from the joint above to the joint below and around the entire limb. Cover the foot with a taped surgical glove and create a tape stirrup so the limb can be suspended for full access during clipping and preparation. Initial preparation occurs in the prep area; final preparation occurs in the surgical suite.
Back surgery: Common for intervertebral disc disease. Clip approximately two vertebral spaces ahead of and behind the surgical site; the patient is positioned in ventral recumbency.
Bovine flank surgery: Clip and prep from the transverse vertebral processes to the ventral abdomen, and from the last rib to the top of the hip. Flank approaches can be performed standing or in lateral recumbency and are useful for procedures such as rumenotomy and caesarean section.
Equine abdominal surgery: Use landmarks similar to those for small-animal abdominal preparation; other procedures are prepared as directed.
Postoperative cleanup
After surgery, move the patient to recovery, dispose of sharps appropriately, begin instrument soaking, and send reusable drapes, wraps, and gowns to laundry or discard them. Once the patient is stable, disinfect horizontal surfaces, lights, tables, mayo stands, and floors, including beneath equipment. Use a broom and mop dedicated only to the surgical suite.
PP ~ Unit 6
Microchipping, tattooing, and proper clipper maintenance provide reliable animal identification and safe grooming equipment use.
Clipper Types and Blade Care
The unit covers regular/shearing clippers and large-animal clippers. Daily blade care includes brushing away hair, using Blade Wash on new and used blades, wiping the blades before use on a patient, and placing a drop of oil between the cutting and comb blades. Do not oil the cutting surface; apply disinfectant or Kool Lube only to the cutting blade.
Clipper Maintenance
Blades must be professionally sharpened. Large-animal clippers need the same routine care plus tension-knob adjustment. Check grease monthly and add or replace it about every 2–4 months, or sooner if the clipper squeals. Inspect the blade latch assembly, felt pads, and carbon brushes; check brushes every 3–6 months and replace when needed.
Microchipping
Microchips use RFID: the scanner actively sends radio waves, while the implanted chip is passive. The chip’s identification number is stored in a centralized recovery database. In dogs and cats, it is placed subcutaneously on the dorsal midline between the shoulder blades.
Scanning and Recovery
When a stray animal arrives, scan for a chip. If found, the scanner displays an alphanumeric code. Use that code to contact the recovery database, which can trace the registered owner information at any time.
Advantages of Microchips
Microchips are easy to place, permanent, non-removable, and do not fade or become lost over time. They can migrate, but they allow rapid access to owner information through the recovery database.
Tattooing
Tattooing is a permanent identification method for both large and small animals and may be required by some breeds. Small-animal tattoo sites include the inside of the pinna, flank, or thigh. Methods include writing tattoos and stamp tattoos; some horse breeds use lip tattoos.
PP ~ Unit14
Waste anesthetic gases are a significant workplace hazard in veterinary settings, but exposure can be substantially reduced through effective scavenging, leak testing, safe techniques, ventilation, and equipment maintenance.
Waste anesthetic gas (WAG)
WAG includes vapors such as isoflurane, sevoflurane, and nitrous oxide that escape into clinic air. All personnel can be affected, so facilities using inhalant anesthetics need a control program.
Health effects
Short-term exposure can cause drowsiness, headache, fatigue, nausea, itching, depression, and irritability; recurring symptoms may signal excessive air concentrations. Long-term exposure has been linked to reproductive effects, liver and kidney disease, bone-marrow abnormalities, and changes in memory or motor performance. Risk increases with dose and duration.
Major exposure sources
Vapors may escape from anesthetic machines, recovering animals’ exhaled breath, leaks, poorly fitting components, mask induction, induction chambers, and liquid-anesthetic spills. Spills are especially hazardous because liquid rapidly evaporates and can be absorbed through intact skin.
Higher-risk conditions and locations
Surgery, dentistry, recovery kennels, and the area near the patient’s mouth or anesthetic machine carry the greatest risk. Exposure rises with longer procedures, higher carrier-gas flow, inadequate scavenging, poor ventilation, and leaky anesthetic techniques.
Exposure limits and ventilation
Recommended limits are 2 ppm for halogenated anesthetics such as isoflurane and sevoflurane, and 25 ppm for nitrous oxide. Good ventilation is essential; the document recommends approximately 15–20 air changes per hour.
Reducing WAG exposure
Use an effective scavenging system, perform machine leak checks, avoid mask induction and induction chambers, use cuffed endotracheal tubes, use the lowest effective oxygen flow, and keep vaporizers off until the patient is intubated and connected. Active scavenging systems use a pump or fan to exhaust gases outdoors and can reduce WAG by 64–94%.
Leak testing and maintenance
Perform a low-pressure leak test before the first use each day, after changing systems, or when another person has used the machine. Check hoses, reservoir bags, seals, valves, the CO2 canister, and vaporizer fittings. Service machines annually or twice yearly and test ET-tube cuffs with every use.
Spill response and respiratory protection
For spills, clear the area, increase ventilation, wear gloves, apron, and an organic-vapor respirator, absorb the liquid with spill material or kitty litter, and dispose of waste in a sealed container outside the clinic. N95 masks provide no protection against WAG; only fit-tested organic-vapor respirators with activated-carbon cartridges are appropriate.
Oxygen cylinders and vaporizers
Store oxygen cylinders upright and secured, keep protective caps on unused tanks, label full and empty cylinders, and transport them by handcart. Open connected tanks slowly while turning the face away from valves and gauges. Keep vaporizers upright; if tipped, allow 12–24 hours for liquid to settle and consider recalibration before use.
Injectable anesthetic safety
Injectable agents can harm staff through skin contact, eye splashes, or ingestion. Potent opioids, cyclohexamines, and alpha-2 agonists can cause severe sedation, respiratory depression, unconsciousness, coma, or death. Use gloves and goggles, ensure trained staff do not work alone with potent opioids, keep reversal agents and first-aid supplies available, and dispose of needles immediately in closed sharps containers.
Equipment-related anesthesia failures
The most common anesthesia-related equipment problems involve breathing circuits (39%), vaporizers (21%), ventilators (17%), and oxygen tanks or lines (11%). Frequent equipment checks help prevent both patient harm and staff exposure.
,