RSM 400 - Comprehensive MRI Physics & Instrumentation Review Guide
Core Principles of MRI Physics
- The Complete MRI Process in Sequence:
- Hydrogen nuclei enter the main static magnetic field ().
- Proton magnetic spins align relative to the field.
- A Net Magnetization Vector (NMV) is established.
- Protons precess around the direction of the external magnetic field.
- A Radiofrequency (RF) pulse is applied at the specific Larmor frequency.
- The Net Magnetization Vector tilts away from the longitudinal plane into the transverse plane ().
- The RF pulse is turned off.
- T1 recovery and T2/T2* decay processes begin simultaneously.
- The emitted magnetic signal is detected by receive coils and spatially encoded.
- The raw signal is processed to reconstruct diagnostic images.
Atomic Structure, Electric Charge, and Electromagnetism Foundations
Atomic Structure:
- The nucleus contains protons () and neutrons ().
- Electrons () orbit surrounding the nucleus.
- Atomic Number (): Represents the total number of protons in the nucleus.
- Atomic Mass (): Represents the combined sum of protons and neutrons ().
- Isotope: Atoms possessing the identical atomic number () but differing in neutron count.
- Ion: An atom with an imbalance between the total number of protons and electrons.
Electrostatics and Coulomb's Law:
- Like electric charges repel one another; opposite electric charges attract.
- Coulomb's Law states that the electrostatic force between two point charges is inversely proportional to the square of the distance separating them:
- Exam Note: If the distance between charges is doubled, the electrostatic force drops to approximately one-fourth () of its original strength.
Moving Electric Charge and Magnetic Fields:
- A stationary electric charge generates purely an electric field.
- A moving electric charge or a spinning charged particle generates a magnetic field. This principle forms a core foundation of MRI, as the hydrogen nucleus consists of a spinning charged proton.
- Exam Note: Moving electric charge magnetic field.
Faraday's Law of Electromagnetic Induction:
- A time-varying magnetic field induces an electromotive force (electric current) within an electrical conductor.
- In MRI, the precessing transverse magnetic signal of hydrogen protons induces a minute electric current within receiver coils, which is digitized to construct the image.
- Exam Note: Changing magnetic field induced electric current (the fundamental mechanism of MR signal reception).
Electromagnetic Wave Characteristics:
- Frequency (): The number of wave cycles per second, measured in Hertz ().
- Wavelength (): The linear distance between two consecutive wave peaks.
- Amplitude: The height or maximum strength of the wave oscillation.
- Frequency and wavelength are inversely proportional.
- All electromagnetic radiation travels through a vacuum at the speed of light: approximately .
Hydrogen, Spin, and Magnetic Moment
Selection of Hydrogen for Clinical MRI:
- Hydrogen is ideal for MR imaging because it contains a single unpaired proton, exists in extreme abundance throughout the human body (primarily in water and fat molecules), and exhibits a high gyromagnetic ratio (), providing strong detectable signal strength.
Spin Quantum Number:
- The single hydrogen proton possesses a spin quantum number of .
- Consequently, when placed in an external magnetic field, it occupies one of two discrete energy orientations: parallel (spin-up) or anti-parallel (spin-down).
- Exam Note: Hydrogen spin quantum number = .
Magnetic Moment and Magnetic Domain:
- Because the hydrogen proton possesses an electric charge and intrinsic spin, it behaves like a microscopic dipole magnet and exhibits a magnetic moment.
- A group of aligned magnetic moments constitutes a magnetic domain.
- In biological tissues outside an external magnetic field, proton magnetic moments are randomly oriented, resulting in a net magnetic vector of zero.
Main Magnetic Field (), Alignment, and Net Magnetization
Static Magnetic Field ():
- represents the powerful, static main magnetic field generated by the MRI scanner.
- When a subject is placed inside , hydrogen magnetic moments tend to align with the axis of the field.
Spin-Up vs. Spin-Down Energy States:
- Spin-Up / Parallel: Magnetic moments align parallel with ; this represents the lower energy state.
- Spin-Down / Anti-Parallel: Magnetic moments align opposite to ; this represents the higher energy state.
- A slight thermal excess of protons exists in the spin-up (parallel) state compared to the spin-down state. This slight population difference produces the Net Magnetization Vector (NMV).
Net Magnetization Vector (NMV):
- The NMV is the vector sum of all individual magnetic moments within a given volume.
- At equilibrium, the NMV lies parallel to the main static magnetic field along the longitudinal axis (-axis), designated as or .
- Exam Note: Vector sum Net Magnetization. Long axis of the patient -axis / .
Units of Magnetic Field Strength:
- Magnetic field intensity is quantified in Tesla () or Gauss ().
- .
- .
- .
- The Earth's natural magnetic field averages approximately .
- Exam Note: .
Magnetic Susceptibility
- Definition:
- Magnetic susceptibility () measures the degree to which a substance becomes magnetized when exposed to an external magnetic field, influencing the homogeneity of the local field.

| Susceptibility Category | Behavior in External Magnetic Field | Behavior After Removal from Field | Examples |
|---|---|---|---|
| Diamagnetic | Weakly repelled () | No residual magnetization remains | Hydrogen, copper, silver |
| Paramagnetic | Weakly attracted () | Minor short-term residual magnetization may remain | Oxygen, manganese, aluminum |
| Superparamagnetic | Strongly attracted | Longer-lasting residual magnetization may remain | Certain magnetic nanoparticles / contrast agents |
| Ferromagnetic | Extremely strongly attracted | Retains permanent magnetization | Iron, cobalt, nickel |
- Exam Note: Gradient Recalled Echo (GRE) pulse sequences are exceptionally sensitive to magnetic susceptibility artifacts because they lack a RF refocusing pulse and do not correct static magnetic field inhomogeneities.
Gyromagnetic Ratio, Precession, and Larmor Frequency
Gyromagnetic Ratio ():
- A fundamental physical constant specific to each nuclear species, defined as the ratio of its magnetic moment to its angular momentum.
- For hydrogen (), the gyromagnetic ratio is approximately .
- Larmor frequency at : Approximately .
- Larmor frequency at : Approximately .
- Exam Note: Higher field strength higher Larmor precessional frequency.
Precession:
- When placed in , proton magnetic moments wobble around the static field axis in a gyroscopic motion called precession.
Larmor Frequency and Resonance:
- The rate of precession is governed by the Larmor Equation:
- Resonance occurs exclusively when an applied external RF pulse oscillates at the exact Larmor precessional frequency of the targeted protons.
- Exam Note: Resonance condition RF frequency = Larmor frequency.
Radiofrequency (RF) Pulse, , and Flip Angle
RF Excitation Field ():
- Applying an RF pulse generates an oscillating magnetic field () perpendicular to .
- When the RF pulse matches the Larmor frequency, protons absorb energy, tipping the NMV away from the longitudinal axis (-axis) into the transverse plane (-plane).
Excitation Pulse:
- A flip angle tips the entire NMV into the transverse plane, establishing maximum transverse magnetization ().
- Exam Note: Maximum flip angle.
Pulse Functions:
- In Spin Echo (SE) sequences: Functions as a refocusing pulse to rephase dephased transverse spins.
- In Inversion Recovery (IR) sequences: Functions as an initial inversion pulse to invert by .
Gradient Echo (GRE) Flip Angles:
- GRE sequences typically employ low flip angles () and rely on gradient magnetic fields rather than RF pulses for spin rephasing.
- Exam Note: GRE flip angle is usually , leading to increased sensitivity to magnetic field inhomogeneities.
Relaxation Processes: T1, T2, and T2*
- T1 Relaxation (Spin-Lattice Relaxation):
- T1 relaxation describes the recovery of longitudinal magnetization () back toward thermal equilibrium.
- Protons give up absorbed thermal energy to the surrounding molecular structure (lattice).

| Elapsed Time Interval | Approximate Longitudinal Magnetization () Recovered |
|---|---|
Exam Note: After an elapsed time of , approximately of longitudinal magnetization has been recovered.
T2 Relaxation (Spin-Spin Relaxation):
- T2 relaxation describes the decay of transverse magnetization () caused by spin dephasing.
- Spin-spin interactions cause adjacent protons to exchange energy, losing phase coherence without losing net energy to the lattice.
- Exam Note: Spin-spin interaction transverse decay / T2 relaxation.
T2* Decay:
- T2* decay reflects total transverse dephasing, combining true T2 spin-spin relaxation with dephasing caused by magnetic field inhomogeneities ( non-uniformity).
- T2* decay occurs much faster than true T2 decay ().
- Exam Note: , and Free Induction Decay (FID) follows T2* decay.
Comparison Between T1 and T2 Times:
- T1 represents longitudinal recovery; T2 represents transverse decay.
- For any given tissue, T1 time is significantly longer than T2 time (`), meaning transverse phase coherence is lost long before longitudinal magnetization fully recovers.
Free Induction Decay (FID) and Signal Detection
Free Induction Decay (FID):
- The immediate decaying sinusoidal signal generated by precessing transverse magnetization right after the excitation RF pulse turns off.
- The signal decays exponentially according to T2* relaxation.
Signal Detection Mechanics:
- According to Faraday's Law, precessing transverse magnetization induces a minute oscillating electrical current within the receiver coil, which is processed into raw MR signal.
- Exam Note: FID is the first signal produced following excitation and decays via T2* relaxation.
Echo Formation and Spin Echo Sequence
- Spin Echo Sequence Mechanics:
- A RF pulse converts longitudinal magnetization into maximum transverse magnetization ().
- Transverse magnetization begins dephasing due to field inhomogeneities and spin-spin interactions.
- At time , a RF refocusing pulse is applied to flip precessing spins around the transverse plane.
- Spins rephase to produce a spin echo, reaching maximum amplitude at Echo Time (TE).

- Sequence Timeline:
- RF excitation pulse maximum created.
- Dephasing period begins.
- RF refocusing pulse applied at .
- Rephasing phase occurs.
- Peak spin echo formed and sampled at TE.
- Exam Note: The inclusion of an RF refocusing pulse is the critical structural difference distinguishing Spin Echo from Gradient Recalled Echo (GRE).
Sequence Parameters: TR and TE
Repetition Time (TR):
- TR is the time interval between consecutive excitation RF pulses applied to the same slice.
- TR controls the degree of longitudinal magnetization recovery permitted between pulses; therefore, TR primarily controls T1 weighting.
- Exam Note: TR controls T1 weighting.
Echo Time (TE):
- TE is the time elapsed from the initial excitation RF pulse to the peak signal collection in the echo.
- TE controls the amount of transverse decay allowed to occur before signal collection; therefore, TE primarily controls T2 weighting.
- Exam Note: TE controls T2 weighting. Extending TE increases T2 weighting.
Image Weighting: T1, T2, and Proton Density (PD)

| Weighting Type | Repetition Time (TR) | Echo Time (TE) | Mnemonics & Memory Tricks |
|---|---|---|---|
| T1-Weighted | Short | Short | SS = Short TR / Short TE |
| T2-Weighted | Long | Long | LL = Long TR / Long TE |
| Proton Density (PD) | Long | Short | LS = Long TR / Short TE |
T1-Weighted Contrast:
- Short TR highlights tissue differences in T1 recovery rates.
- Short TE prevents T2 decay differences from obscuring contrast.
- Fat appears hyperintense (bright).
- CSF / fluid appears hypointense (dark).
- Provides high contrast for anatomical structure visual analysis.
- Exam Note: ; Fat is bright on T1.
T2-Weighted Contrast:
- Long TR minimizes T1 contrast contribution (suppresses T1 differences).
- Long TE allows substantial T2 spin-spin decay differences between tissues to develop.
- Water / CSF appears hyperintense (bright).
- Edema and pathological fluid collections appear hyperintense (bright).
- Essential for detecting pathology.
- Exam Note: ; Water/fluid is bright on T2.
Proton Density (PD) Weighting:
- Reflects relative differences in mobile hydrogen proton concentration () per tissue volume.
- Uses Long TR to minimize T1 effects and Short TE to minimize T2 effects.
- Exam Note: .
Dual Echo Spin Echo Sequences:
- Employs a single Long TR combined with two distinct Echo Times (TE).
- First echo (Short TE) yields a Proton Density-weighted image.
- Second echo (Long TE) yields a T2-weighted image.
- Exam Note: Dual echo sequence combined PD + T2 images.
Tissue Relaxation Characteristics and Signal Intensity
- Molecular Environment and Mobility:
- Different tissue structures exhibit unique T1 and T2 relaxation times determined by molecular size, tumbling rate, and bound vs. free water status.

| Tissue Type | Approximate T1 Time () | Approximate T2 Time () | General Visual Appearance & Signal Characteristics |
|---|---|---|---|
| Fat | Short T1 hyperintense (bright) on T1 | ||
| White Matter (WM) | Relatively short T2 | ||
| Gray Matter (GM) | Brighter than White Matter on T2 | ||
| Edema | Long T2 hyperintense (bright) on T2 | ||
| CSF | Very long T1 & T2 extremely bright on T2 |
Why Fat is Bright on T1-Weighted Images:
- Fat possesses a short T1 relaxation time due to its slow tumbling rate matching the Larmor frequency, enabling rapid longitudinal recovery ().
- When a short TR is applied, fat has already recovered substantial longitudinal magnetization, yielding high signal intensity on subsequent pulses.
Why Water/CSF is Dark on T1-Weighted Images:
- Water molecules move rapidly, causing a poor match with the Larmor frequency and resulting in a long T1 relaxation time.
- At a short TR, water longitudinal magnetization recovers poorly, yielding low signal intensity.
Why Water/CSF is Bright on T2-Weighted Images:
- Water has a long T2 relaxation time because small molecules tumble quickly, minimizing internal magnetic field fluctuations and delaying spin dephasing.
- Even at a long TE, water retains substantial transverse coherence (), resulting in high signal output.
Phase, Dephasing, and Rephasing
Phase:
- Refers to the specific angular position of magnetic spins precessing relative to one another at a given time point.
In-Phase vs. Dephasing:
- When proton spins point in the exact same direction, they are in-phase, generating a large transverse vector ().
- When spins fan out at different rates, dephasing occurs, rapidly diminishing net transverse magnetization and signal strength.
Rephasing / Refocusing:
- The process of bringing precessing spins back into phase coherence to rebuild a measurable signal echo.
- Exam Note: Phase = angular position of magnetic spins relative to time.
Gradient System and Spatial Encoding
Gradient Coils ():
- Magnetic gradient coils impose controlled, linear variations on the static magnetic field ().
- Their primary function is spatial encoding, ensuring signals can be attributed to specific physical spatial locations within the patient.
Three Dimensions of Spatial Encoding:
- Slice selection gradient ( in conventional axial imaging).
- Phase encoding gradient ().
- Frequency / readout encoding gradient ().
Mechanism:
- Gradients alter precessional frequency and phase of protons as a function of location, labeling their position.
- Combining three orthogonal gradients allows arbitrary slice plane orientation (axial, sagittal, coronal, or oblique).
- Exam Note: Spatial encoding gradient field system.
Main Hardware Components of an MRI Scanner
Concentric Hardware Layers (Outer to Inner):
- Main Static Field Coil ().
- Gradient Coil System ().
- RF Body Coil / Transceiver (closest to patient).
Superconducting Magnets:
- Clinical MRI systems primarily use superconducting main magnets constructed from niobium-titanium () alloy wire.
- Coils are submerged in liquid helium () to maintain superconductivity (zero electrical resistance).
- Zero resistance allows continuous, high-ampere electric currents to generate powerful, stable static magnetic fields.
Body Coil and Surface Receiver Coils:
- The integrated body coil can transmit RF and receive signal, but dedicated local receiver coils are placed near target body parts to maximize Signal-to-Noise Ratio (SNR).
- The transmitted RF field must match the precessional Larmor frequency.
Pre-Scan Calibration Automated Steps:
- Center frequency tuning: Finds the exact Larmor frequency for the targeted tissue region.
- RF power adjustment: Calculates precise transmit power needed to achieve desired flip angles.
- Specific Absorption Rate (SAR) monitoring: Estimates RF power absorption to prevent tissue overheating.
Signal-to-Noise Ratio (SNR) Optimization
Signal-to-Noise Ratio (SNR):
- Defined as the ratio of meaningful biological signal intensity to background noise.
- Higher SNR yields clearer visual resolution and improved diagnostic accuracy.
Key Factors Influencing SNR:
- Proton Density (): Higher hydrogen proton density produces greater signal.
- TR Length: Longer TR permits greater longitudinal recovery (), yielding higher signal.
- Sequence Choice: Conventional Spin Echo sequences generate higher SNR and exhibit lower susceptibility artifact sensitivity compared to Gradient Echo sequences.
Gradient Recalled Echo (GRE) Key Principles
Mechanism:
- GRE sequences employ gradient field reversals instead of RF refocusing pulses to generate echoes.
Key Characteristics:
- Flip angles are typically low ().
- No RF refocusing pulse is applied.
- Faster acquisition speeds.
- Substantially more sensitive to static field inhomogeneities ( non-uniformities) and susceptibility artifacts.
- Demonstrates pronounced T2* decay effects.
- Exam Note: If asked which pulse sequence explicitly emphasizes magnetic susceptibility artifacts Gradient Recalled Echo (GRE).
Summary of High-Yield Exam Review Points
- Susceptibility artifact GRE sequence.
- Quantum spin number of hydrogen = .
- After approximately longitudinal magnetization recovered.
- Spin-spin interaction transverse decay / T2 relaxation.
- Free Induction Decay (FID) initial signal emitted after excitation, decaying via T2* relaxation.
- Resonance condition applied RF pulse must oscillate at the exact Larmor precessional frequency.
- Maximum transverse magnetization () produced by a flip angle.
- RF pulse function refocusing in Spin Echo; inversion in Inversion Recovery.
- GRE sequences commonly use flip angles
- Earth's average magnetic field strength .
- Gyromagnetic ratio of hydrogen .
- .
- Vector sum of individual magnetic moments Net Magnetization Vector (NMV).
- TR controls T1 weighting; TE controls T2 weighting.
- T1 weighting Short TR + Short TE.
- T2 weighting Long TR + Long TE.
- Proton Density (PD) weighting Long TR + Short TE.
- Increasing TE increases T2 weighting.
- T2 weighting highlights fluid, water, and edema.
- Dual echo sequence yields combined PD + T2 images.
- Spatial encoding achieved via the gradient system.
- Long axis of the patient -axis / M_z$.\n* Moving electric charge \rightarrow generates a magnetic field.\n\n# Common Exam Traps and Pitfalls\n\n* **Do Not Confuse Relaxation Directions**:\n * T1 = Longitudinal recovery (M_z / spin-lattice).\n * T2 = Transverse decay (M_{xy} / spin-spin).\n* **FID Decay Governing Mechanism**:\n * FID decay is governed by T2* decay, NOT pure T2 decay.\n* **Parameter Control Traps**:\n * Long TE emphasizes T2 weighting, NOT T1 weighting.\n * Short TR creates T1 contrast; Long TR suppresses T1 tissue contrast differences.\n* **PD vs. T1 Parameter Distinctions**:\n * Do not mistake PD for T1. PD weighting requires a Long TR combined with a Short TE.\n* **Spin Echo vs. GRE Distinctions**:\n * Spin Echo uses a 180^\circ180^\circ RF pulse.\n* **Energy Level Orientations**:\n * Parallel / Spin-Up = Lower energy state.\n * Anti-Parallel / Spin-Down = Higher energy state.\n* **Field Strength Conversions**:\n * 1\,\text{T} = 10,000\,\text{G}3\,\text{T} = 30,000\,\text{G}.\n* **Maximum Transverse Magnetization**:\n * Achieved exclusively at a 90^\circ flip angle.\n* **T2* vs. T2 Duration**:\n * \text{T2}^* < \text{T2} because field inhomogeneities accelerate dephasing.\n\n# Final Exam Rapid Review Cheat Sheet\n\n\n\n| Prompt / Cue Seen on Exam | Correct Answer / Matching Concept |\n| :--- | :--- |\n| TR | T1 |\n| TE | T2 |\n| Short TR + Short TE | T1-weighted |\n| Long TR + Long TE | T2-weighted |\n| Long TR + Short TE | PD-weighted |\n| Fluid / edema bright | T2 |\n| Fat bright | T1 |\n| First signal after RF | FID |\n| FID decay | T2* |\n| Spin-spin interaction | T2 |\n| Spin-lattice relaxation | T1 |\n| \text{T1} \times 395\% recovery |\n| 90^\circM_{xy} |\n| 180^\circ in Spin Echo | Refocusing pulse |\n| \text{RF} = \text{Larmor} | Resonance |\n| 3\,\text{T}30,000\,\text{Gauss} |\n| Hydrogen spin quantum number | 1/2$$ | | Spatial encoding | Gradients | | Susceptibility emphasized | GRE | | Dual echo | PD + T2 |