PT CARE Unit 1 lecture 8/18/25

Basic X-ray System Components

  • X-ray tube: produces X rays; housed in a lead-lined glass/tube; powered by electricity; X rays are a form of energy that must be generated by the machine.

  • Collimator: controls and shapes the beam (rectangles or squares); uses a light to visualize the beam path since X rays are invisible.

  • Radiographic table: patient support; typically height-adjustable; four-way floating to move in multiple directions; radiolucent to avoid blocking X rays.

  • Bucky: the image receptor assembly (upright bucky for chest, table bucky); can be built-in or use detachable digital plates; modern plates are wireless to computer.

  • Tube supports: ceiling-mounted (common in hospitals) or floor-mounted (older/office setups); overhead crane allows movement across room.

  • Generator: provides the high voltage electricity to produce X rays.

  • Control console: interface to set exposure parameters; newer systems may use voice activation or touch controls; includes audible/visual cues when beam is on.

  • Image receptor options: historical film, now digital plates; typical plate size is 14'' \times 17'' (35 cm x 43 cm) and can be placed portrait or landscape.

  • Alignment concept: radiation field must align with image receptor; improper alignment wastes dose and degrades image.

  • Terminology readiness: expect to learn terms for positions, movements, and safety.

Beam Generation and Imaging Chain

  • X-ray production requires electricity; X rays exit the tube, pass through the collimator, then reach the patient and finally the image receptor.

  • Collimation confines the beam to the anatomy of interest to minimize patient exposure.

  • The image receptor captures transmitted X rays to form the image; beam should not extend beyond the receptor.

Collimation and Field Size

  • Field shapes: rectangles or squares; rounded shapes are not supported by typical collimators.

  • Field size indicators show dimensions in the machine readout; distance indicators show the patient-to-receptor setup in cm.

  • Automatic collimation (PBL) vs manual collimation; PBL automatically adjusts field size to fit the plate size.

  • Field size practice: field should be appropriate for the part (e.g., chest vs finger) and ideally not larger than the image receptor.

  • Typical full-size field: 14'' \times 17'' (or 35\,\text{cm} \times 43\,\text{cm}); fields can be reduced for smaller anatomy.

Image Receptors, Bucky, and Image Workflow

  • Upright bucky for chest X-rays; table bucky under the patient for table exams.

  • Digital plates: wireless transfer to computer; older systems required wired connections.

  • Radiolucent table materials (carbon-graphite fiber) allow X rays to pass with minimal attenuation.

  • Radiopaque materials (e.g., metal markers) show up white on images.

  • Alignment: ensure tube, patient, and image receptor are in proper alignment for accurate imaging.

X-ray Table and Safety

  • Tables are usually four-way floating to ease positioning; most are motorized or foot-pedal controlled.

  • Trendelenburg: head lower than feet; Fowler: head higher than feet.

  • Table safety: lock wheels, use side rails, avoid leaving a patient unattended on the table.

  • Transfer safety: use slide boards when moving patients; ensure patient hands are visible to prevent injury.

  • Radiolucent skin and soft tissues appear relatively faint on X-ray; bones and metals are more radiopaque (white).

Tube Movement and Geometry

  • Longitudinal (along the table’s long axis) movement; Transverse/Lateral (short axis) movement.

  • Vertical movement: raise/lower the tube height.

  • Angulation: cephalad (toward head) or caudal (toward feet); some exams use transverse angulation.

  • Detent: magnetic lock that latches positions; some systems require a click to lock.

  • All movement and positioning should maintain alignment among tube, patient, and receptor.

Source-to-Image Distance (SID) and Vertical Positioning

  • SID: distance from the X-ray source to the image receptor; affects magnification and exposure.

  • Chest radiographs commonly at 72'' (approx. 180\,\text{cm}); extremities at 40'' (approx. 100\,\text{cm}).

  • Vertical positioning defines how the tube is placed relative to the patient.

Exposure Fundamentals

  • Exposure consists of Quantity (mass) and Quality (penetration).

  • MA (milliamperes) controls the amount of X rays produced; time controls how long exposure lasts.

  • MAS (MA × time) gives the total radiation dose delivered.

  • KV (kilovoltage) controls beam quality and penetrating power.

  • Formula relation: MAS = MA × time; higher KV increases penetration, not necessarily dose.

  • Anatomic programming and AEC (automatic exposure control) help estimate the needed exposure and stop when images are adequate.

  • Not all exams use AEC; some require manual setting of MAS and KV based on body part and patient size.

Automatic Exposure Control (AEC) and Technique

  • AEC uses detectors to terminate exposure when sufficient signal is reached.

  • Anatomical programming provides starting MAS and KV values; adjustments may be needed for size, projection, or pathology.

  • You must still position the patient accurately; poor positioning can’t be fully corrected by exposure settings.

Portable (Mobile) X-ray and Fluoroscopy

  • Portables/mobile units used when the patient cannot be moved; battery-powered with charging.

  • Wireless or wired image receptors; images can be viewed on a screen at the bedside or on a central system.

  • Fluoroscopy: live X-ray (dynamic imaging) with an image intensifier and C-arm or U-arm; used in interventional radiology and some diagnostic procedures.

  • C-arm: portable fluoroscopic system that can move around the patient; essential in OR for procedures like cholangiograms, pacemaker placements, and orthopedic work.

Interventional Radiology and Safety

  • Interventional radiography uses live imaging to guide procedures (angioplasties, stents, vessel work).

  • Safety: avoid banging the tube or collimator into patient; keep alignment; prevent excessive tube rotation due to high voltage cables.

  • Always secure the patient with rails, locks, and proper transfer techniques; never leave a patient unattended on the table.

  • Kill switchs and emergency controls exist for safety; be familiar with their location and purpose.

Practice and Mindset

  • Imaging is a combination of physics, technique, and patient care.

  • You are problem-solving: choose appropriate MAS and KV, ensure proper SID and alignment, and use AEC/technique as needed.

  • Real-world workflow includes moving equipment, positioning patients, and coordinating with teams; safe, precise, and efficient practice is essential.

Quick Reference Terms (for review)

  • Longitudinal: movement along the table’s long axis; Transverse/Lateral: along the short axis.

  • Cephalad: toward head; Caudal: toward feet.

  • Trendelenburg: head lower than feet; Fowler: head higher than feet.

  • SID: Source-to-Image Distance; typical chest: 72''\ (180\,\text{cm}); extremities: 40''\ (100\,\text{cm}).

  • MAS: \text{MAS} = \text{MA} \times t; KV: kilovoltage; MA: milliamperes.

  • Field size: rectangle/square; maximum approx. 17''\times17'' or 14''\times17''; typical default 10''\times10''.

  • Image receptor: film vs digital plate; bucky alignment with tube is essential for proper exposure.