Pulmonary Function Testing for Respiratory Therapists

Overview of Pulmonary Function Tests (PFTs)

  • Definition: PFTs are diagnostic measurements aimed at evaluating lung function.

  • Normal Lung Function Determination: Normal function is context-dependent, influenced by:

    • Age: Lung function typically declines with age.

    • Gender: Males generally have larger lung volumes compared to females.

    • Height: Taller individuals usually have larger lung volumes.

  • Criteria for Normal Function: In general, lung function is considered normal if it is ≥ 80% of the predicted value.

Objectives of PFTs

  1. Differentiation: Distinguishing between obstructive and restrictive lung diseases.

  2. Severity Assessment: Determining the severity of lung diseases.

  3. Monitoring: Tracking the progress of chronic respiratory conditions.

Lung Volumes

  • Inspiratory Reserve Volume (IRV): The amount of air that can be inhaled after a normal tidal inhalation.

  • Tidal Volume (VT): The volume of air inhaled or exhaled during normal breathing.

  • Expiratory Reserve Volume (ERV): The volume of air that can be forced out after normal exhalation.

  • Residual Volume (RV): The amount of air that remains in the lungs after a forced exhalation (cannot be exhaled).

Lung Capacities (Sum of Two or More Volumes)

  • Inspiratory Capacity (IC): The maximum volume of air that can be inhaled after normal expiration, calculated as:

    • IC=IRV+VTIC = IRV + VT

  • Vital Capacity (VC): The total amount of air exhaled after a maximum inhalation, calculated as:

    • VC=IRV+VT+ERVVC = IRV + VT + ERV

  • Functional Residual Capacity (FRC): The volume of air remaining after a normal tidal exhalation:

    • FRC=ERV+RVFRC = ERV + RV

    • Note: Decreased FRC can lead to decreased oxygenation.

  • Total Lung Capacity (TLC): The maximum amount of air accommodated by the lungs, calculated as:

    • TLC=IRV+VT+ERV+RVTLC = IRV + VT + ERV + RV

Measurement of Lung Volumes

Method: Spirometry
  • Spirometry: The primary technique for measuring lung volumes by assessing airflow.

    • Basic Equations:

    • Volume = Flow × Time

    • Flow = Volume / Time

    • Spirometer: Device used to measure exhaled lung volumes.

    • Spirogram: A graph plotting volume against time.

Limitations of Simple Spirometry
  • Simple spirometry cannot measure lung volumes that cannot be exhaled (e.g., RV, FRC, and TLC).

  • These volumes can be assessed using:

    • Body Plethysmography: Most accurate, utilizes Boyle’s Law.

    • Nitrogen Washout.

    • Helium Dilution.

Vital Capacity (VC) Measurement

  • VC Measurement Methods:

    • Slow Vital Capacity (SVC): Used when volume alone is a concern. May indicate:

    • Restrictive Diseases: Decreased volumes.

    • Obstructive Diseases: Increased volumes.

      • RV/TLC Ratio: Increased, indicating hyperinflation/air trapping.

    • Forced Vital Capacity (FVC): Used when evaluating flow rates.

    • Obstructive Disorders: Characterized by decreased flows due to exhalation obstruction.

    • General Presentation: Obstructive lung disorders exhibit high lung volumes, meaning they can inhale air but have difficulty exhaling it.

Measurement of Flows

Pulmonary Dynamics
  • Forced Vital Capacity (FVC): Measures total air exhaled forcefully.

  • Forced Expiratory Volume (FEV) Timed: Measures the volume exhaled in the first second, second, etc.

  • FEV1/FVC Ratio: This compares the amount exhaled in the first second to the total FVC:

    • Normal and Obstructive Conditions: FEV1 is decreased in both but for different reasons:

    • In obstructive disorders, due to limited airflow.

    • In restrictive disorders, due to already low lung volumes.

Patient Instructions for FVC Measurement

  • Direct the patient to:

    • Take a deep breath and blow all air out as hard and fast as possible.

Repeatability and Reliability of Measurements

  • Criteria for Repeatability:

    • The difference between the two largest FVC or FEV1 values must be < 0.150 L.

    • Example of Reporting: If the values are 3.5 L, 3.0 L, and 2.8 L, they are inconsistent and cannot be reported since the highest values (3.5 and 3.0) are not within the allowed range.

FEV1/FVC Data Interpretation

  • Used to differentiate obstructive and restrictive lung conditions.

  • Normal or High in Restrictive Disorders: Indicates adequate flow performance despite volume issues.

  • FEV1/FVC Ratio < 70% indicates obstructive lung disease.

  • Peak Expiratory Flow Rate (PEFR):

    • Indicates maximal achievable expiratory flow during FVC maneuver, useful in assessing airflow limitation in asthma, but may underreport limitations in COPD.

Flow-Volume Loop (F-V Loop)

Normal F-V Loop
  • In normal lungs, airway diameter decreases as air is exhaled, shown as a gradual decrease in flow.

Obstructive F-V Loop
  • Affected lungs demonstrate significant flow decline post-PEFR and exhibit a scooped appearance due to airflow limitations.

Effort Dependence of Measurements

  • PEFR: Depends on muscular effort exerted during forced expiration.

  • FEF 25-75% Measurement: An effort-independent measurement that assesses flow rates after maximum flow has been achieved, unaffected by additional effort.

Fraction of Exhaled Nitric Oxide (FENO)

  • Used in asthma diagnostics.

  • Elevated FENO values support asthma diagnosis when eosinophilic inflammation is present:

    • Normal Value: < 25 parts per billion.

Dynamic Compression Explained

Passive Expiration
  • Negative pressure outside airways keeps them open during exhalation.

Forced Expiration and Dynamic Compression
  • In forced expirations, pleural pressures become positive and can lead to airway collapse.

  • Equal Pressure Point: The point at which airway pressure equals the surrounding pressure, inhibiting airflow.

  • Increased effort during exhalation can lead to decreased airway radius and flow.

Implications of Dynamic Compression in Obstructive Disease

  • Patients with obstructive disease (e.g., COPD) experience airway collapse and increased airway resistance during forced exhalation, leading to air trapping.

  • Pursed-Lip Breathing: Technique to counteract dynamic compression by maintaining positive pressure in airways and extending exhalation time.

Calibration and Quality Control for PFT Equipment

  • Compliance with Standards: All PFT equipment should meet ATS - ERS standards.

  • Calibration Process:

    • Use a large volume syringe (3.0 liters) for volume calibration.

    • Accuracy Requirement: Must be within +/- 3.5%, so acceptable volume range is between 2.895 L and 3.105 L.

    • Flow Calibration: Conducted using a rotameter, ensuring a range of 2 to 12 L/sec.