Protein Quantitation Methods
Protein Quantitation
Overview of Current Methods
Protein quantitation techniques include:
Amino Acid Analysis
Most accurate but requires sophisticated instruments.
UV Spectrophotometry
Colorimetric Methods
Mass Spectrometry
UV Spectrophotometry
Principle: Measures absorbance of aromatic amino acids at 280 nm to quantify protein concentration.
Amino Acids Used:
Tryptophan
Tyrosine
Cysteine
Advantages:
Simple method that requires small sample volume.
No standard curve needed if molar absorption coefficient is known.
Disadvantages:
Requires pure proteins and is incompatible with contaminants like nucleic acids with similar absorption spectrum.
Absorption coefficient must be pre-determined.
Colorimetric Assays
Function: Determines protein concentration in solutions through optical density measurement.
Common Assays (Dye-binding):
Lowry Assay
BCA Assay (Bicinchoninic Acid)
Bradford Assay
Process:
Uses a standard curve of a reference protein (e.g., BSA) to determine concentration.
Beer-Lambert Law
Relates optical density (A) to concentration (c) using the equation:
A (optical density) = εlc
Where:
E = Absorption coefficient
c = Concentration of the compound
l = Optical path length
Proteins for Standard Curves
Pure proteins
Known concentration measured by an independent method
Broad linear dynamic range
Fitting linear or nonlinear regression models to the quantitative data
Principles of Standard Curve Assays
Identically treated samples = directly comparable. Protein quantity = only cause for difference in final absorbance/fluorescence
Validation: If unknown and standard samples have the same absorbance, they contain the same concentration.
Concentration results expressed in the same units as standards to make standard curve (e.g., mg/mL).
Dilution factor is considered post-measurement.
Densitometry Analysis
Usage: Useful when proteins have low UV response, or purity is unclear. Some colorimetric assays have low sensitivity and require a lot of sample.
Measuring Intensity: Optical density: intensity of staining of bands in gel/degree of darkness of X-ray film
Measures degree of darkness as a fcuntion of light transmission.
Equation: OD = Log_{10} (1/Transmittance) = E*c*l
Transmittance: fraction of light not absorbed by sample
Requires a reference standard curve for quantitation.
Benefits of Quantification after Electrophoresis
Allows evaluation of purity and yield of individual proteins in complex mixtures.
Quantification can be relative, particularly if compared to a standard curve based on a different protein.
Standard Curves in Electrophoresis
Created by plotting optical density of protein standards across increasing concentrations.
Data can fit models such as linear, hyperbolic, quadratic, or polynomial for smooth curves.
Imaging Systems
Several systems for capturing signals post-gel electrophoresis exist, including:
Chemiluminescence (after Western blotting)
Transmitted visible light through gel staining
Detection: Converts photon energy to electrical signals using PMT or CCD arrays.
Digitalization of Images
Converts the continuous analog signal to a digital signal.
Predetermined discrete intensity levels.
Number of intensity levels defines digital resolution; greater levels improve sensitivity.
Image Depth and Bit Depth
Image depth refers to pixel intensity values; defined by bits = describe total number of gray shades visible.
Example:
8 bits: 256 shades (2^8)
16 bits: 65536 shades (2^16)
Human eye perceives 6 bits.
Chip with more bits = distinguish smaller differences in intensities
Dynamic Range
Dynamic range: range of band intensities measurable in one instance relating to system noise.
A wide dynamic range = high sensitivity (detection of faint bands) without signal saturation (strong signal bands)
Linear Dynamic Range: Signal intensities are directly related to the source amount.
CCD Camera-Based Systems
Composed of illumination and lens assembly to focus images on a CCD array.
Sensitive to light, temperature, high energy radiation.
Noise from dark current from thermal energy can have strong effect on performance
Designed for single field imaging but may require cooling (very $$$) to reduce noise —> improve sensitivity and dynamic range.
Limitations: CCD chips are small (~1cm²), image much larger —> long optical path is required to rpoject entire image area onto chip. = decrease light collection efficiency, requires long integration times to collect sufficient signal
Digital Image Scanners
Maximize light collection to increase sensitivity: sensor of scanner very close to image —> increase light collection 100-400 fold in a very short time.
Shorter exposure times reduce dark noise, and usable at room temperature with cost-efficient sensors.
Summary of Signal Intensity and Source Proximity
The closer the light source to the sensor, the higher the signal intensity collected (Inverse Square Law).
Example:
C-DiGit Blot Scanner (~1mm source) vs. CCD (~500mm).
The intensity collected decreases with the square of distance from the source.