1/121
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
Reference Range
represents expected values for a healthy individual based on specific population served by the lab
Diagnostic Sensitivity
probability of detecting ALL patients with target disease, negative rules out. no false negatives
Analytical Sensitvity
how sensitive a test is in the analyzer
Diagnostic Specificity
probability of detecting ONLY target disease, positive rules in. No false positives
Analytical Specificity
how specific the analyzer is when running a test
Negative Prediction Value
probability that a negative result means a patient does not have the disease
Positive Prediction Value
probability that a positive result means the patient has the disease
PPV/NPV effect with rare conditions
PPV is lower, NPV is higher
PPV/NPV effect with common conditions
PPV is higher, NPV is lower
Specificity Formula
S = TN/TN+FP
PPV Formula
PV = TP/TP+FP
NPV Formula
PV = TN/TN+FP
Sensitivity Formula
S = TP/TP+FN
Quality Assurance Purpose
assess effectiveness of labs policies and procedures to assure accurate, reliable, and prompt reporting of results
Calibration
standard is used to establish analyzer detection of known value
When is calibration done?
after instrument installation, new lot of reagent, regular intervals, recalibrate after QC trend/shift
Analytical Measurement Range (AMR)
range of results in which instrument response and specimen value is linear, limitation of analyzer/reagent
Clinically Reportable Range (CRR)
may be different from AMR if linear range is extended with dilution
Quality Control
monitor and evaluation of quality of each analysis method, measures accuracy and precision
QC Material
matrix similar to actual specimen, ideally stable with little vial to vial variation
Lyophilized material
freeze dried, reconstitution is required and can be source of error
Assayed
already been tested, very specific target and range, expensive
Unassayed
target and range must be established, broader range requiring techs to run 30x over 30 days, less expensive
Standard Deviation
how widespread values are around the mean, higher is less precise
Coefficient of Variation Formula
SD/mean x 100
Coefficient of Variation
relative standard deviation expressed as %, compares methods
Trend
gradual increase or decrease, can be caused by reagent deterioration or light deterioration
Shift
abrupt increase or decrease, can be caused by change in calibration, reagent change, or analyzer issue, analyzer issue
Random Error
operator technique, pipette problems, incorrect reconstitution
Systematic Error
change of lot, deterioration, improper alignment of sample or reagnet pipette
Westgard Multirules
6 rules for evaluating QC data
1-2s
1 point outside 2SD. serves as warning, may indicate random error
1-3s
1 point outside 3SD. resolve error before releasing patient results, may indicate random error
R-4s
1 point outside +2SD and -2SD within same run. Resolve before releasing patient results, may indicate random error
2-2s
2 points outside 2SD. resolve error before releasing patient results, may indicate systematic error
4-1s
4 points outside 1SD. Resolve error before releasing patient results, may indicate systematic error
10X
10 points on one side of mean. Resolve error before releasing patient results, may indicate systematic error
Random Error troubleshooting
re-assay fresh aliquot, open/reconstitute new vial of control
Systematic Error troubleshooting
replace reagent, instrument/alarm issues
Delta Flag
LIS compares patients current results to previous and finds change greater than threshold
Proficiency Testing
surveys or external quality controls, required by CLIA, tested by CAP and API. Uses simulated patient samples to confirm analyzers are working in universal manner
Wavelength of visible light
400nm to 700nm
Spectrophotometry and Turbidimetry measures:
light absorbed by analyte
Fluorometry and Chemiluminescence measure:
light emitted by analyte
Nephelometry measures:
light scattered by analyte
Light Source
provides light of appropriate wavelength
Monochromator
selects a narrow wavelength bandpass to improve specificity of target analyte
Diffraction Grating
reflective surface with fine grooves that seperates light
Sample Holder
flow cell or cuvette
Photodetector
converts light into electrical signal
Phototube
photodetector that is largely obsolete
Photomultiplier Tube (PMT)
uses dynodes to amplify small signal
Photodiode Array (PDA)
measures multiple wavelength at once
Signal processing
converts electrical signal to result
Reflection Photometry (Light source, monochromator, sample holder, detector, measurement)
LED, filters, test strip/pad, photodiode, reflected light intensity
Spectrophotometry (Light source, monochromator, sample holder, detector, measurement)
Tungsten-Halogen/Deuterium, diffraction grating, cuvette, photodiode/PDA, decrease in transmitted light
Turbidmetry (Light source, monochromator, sample holder, detector, measurement)
Tungsten-Halogen/Deuterium/Xenon, diffraction grating, cuvette, photodiode, decrease in transmitted light
Nephelometry (Light source, monochromator, sample holder, detector, measurement)
laser, filters, cuvette, photodiode/photomultiplier tube, scattered light
Fluorometry (Light source, monochromator, sample holder, detector, measurement)
LED/laser, filters, cuvette, photomultiplier tube, emitted light
Chemiluminescence (Light source, monochromator, sample holder, detector, measurement)
none, filters, reaction cuvette (magnets), photomultiplier tube, light emitted during chemical reaction
Deuterium
Light source for UV
Tungsten/Tungsten-halogen
Light source for visible light
Xenon
Light source for UV and visible light
Ion-Selective Electrode (ISE)
measures for electrolytes using potential voltage between two electrodes
Electrolytes
sodium, potassium, and chloride
Peak position
id
Peak height/area
concentration
Electrophoresis
use of an electric field to separate charges based on charge and size
Anode
Positive
Cathode
Negative
Components of Electrophoresis
electric field, support medium, buffer, sample, detecting system
Immunofixation
antibodies used to identify protein present in bands
Isoelectric Focusing
pH gradient seperates proteins by their isoelectric point
Chromatography
consists of mobile phase and stationary phase. Stationary compounds retain longer, like dissolves like
Retention Time
Property of analyte, how long to elute off
Resolution
property of system, how well does it separate compounds in sample
Absorption
liquid mobile and solid surface stationary
Partition
polar (aqueous) liquid phase and nonpolar (organic) liquid phase
Steric Exclusion
liquid mobile phase and solid stationary phase, based on size and shape
Ion Exchange
opposite charge liquid mobile phase and solid stationary phase
Thin Layer Chromatography (TLC)
mostly obselete, uses Retention factor (Rf)
Retention Factor
Rf = distance of component/total distance of solvent
High-Performance liquid chromatography
therapeutic drug monitoring and toxicology
Gas Chromatography
therapeutic drug monitoring and toxicology
Mass Spectrometry
identify and quantify molecules in a sample based on mass to charge ratio
MALDI
ionized by source of ions for mass spectrometry
Immunoassay
the binding of antibody to antigen for the specific and sensitive detection of an analyte
Competitive Assay Binding
labeled reagent competes with patient analyte for antibody binding sites, signal is inversely proportional to analyte concetration
Non-competitive Assay Binding
labeled reagent used to bind to and directly detect patient analyte. signal is directly proportional to analyte concentration
Heterogenous
requires separation of immune complexes from free antibodies and antigen to measure labeled complex
Homogenous
does not require separation, unbound label will not interfere
Fluorescence Polarization Immunoassay (FPIA)
homogenous competitive. polarized light excites the fluorescent label. Obsolete, less sensitive than CLIA/EIA. Therapeutic drug monitoring and drug testing
ELISA
solid phase heterogenous assay. detection by spectrophotometry. tests infectious disease and autoimmune antibodies
Horseradish peroixdase and tetramethylbenzidine
turns blue then yellow after stop
alkaline phosphatase and p-nitrophenyl phosphate
turns yellow
Chemiluminescent Immunoassay (CLIA)
heterogenous asasy. luminescent labels emit a photon of light as result of chemical reaction
HAMA
human anti-mouse anitbodies in patient sample, exposure to mice
Heterophile antibodies
weak general antibodies that bind to assay
High Dose Hook Effect
false negative/decreased result due to high amounts of analyte
Prozone
antibody excess causing high dose hook effect